9.5. Maintenance Schedule
9.5.1. User Routine Checks Items
The user of the equipment must follow the user routine check items in the table below. It is for the operator
of the equipment to ensure that all checks have been satisfactorily met before attempting to use the
equipment.
It is the operator’s responsibility to provide these checks:
Availability and integrity
Ensure correct functioning
Indicators and buzzer test
Ensure correct functioning
Ensure correct functioning
Ensure correct functioning
Ensure correct functioning
Inspect for kinks and / or cracks
Handswitch and
footswitch
German X-Ray ordinance Roentgenverordnung (RöV) §16 and Qualitätssicherungs-Richtlinie (QS-RL) 3.1.3.8
require the SMPTE test pattern to be displayed on the system monitors for testing per Feb 2001 DIN V
6868-57. Clicking “SMPTE Test” button on Maintenance Setup screen, the operator can perform SMPTE test
on the monitors.
Before using the equipment, verify that system operation is proper according to the check
items above. System that is not fully functional may result in delays in medical treatment.