Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 9-9
9.5. Maintenance Schedule
9.5.1. User Routine Checks Items
The user of the equipment must follow the user routine check items in the table below. It is for the operator
of the equipment to ensure that all checks have been satisfactorily met before attempting to use the
equipment.
It is the operators responsibility to provide these checks:
Item
Description
Frequency
All accessories
Availability and integrity
Daily
Power supply check
Ensure correct functioning
Daily
Indicators and buzzer test
Ensure correct functioning
Daily
X-Ray control check
Ensure correct functioning
Daily
Vertical movement check
Ensure correct functioning
Daily
Brakes and castors
Ensure correct functioning
Daily
Cabling
Inspect for kinks and / or cracks
Daily
Inscriptions and labels
Check for legibility
Annually
Handswitch and
footswitch
Check functionality
Daily
German X-Ray ordinance Roentgenverordnung (RöV) §16 and Qualitätssicherungs-Richtlinie (QS-RL) 3.1.3.8
require the SMPTE test pattern to be displayed on the system monitors for testing per Feb 2001 DIN V
6868-57. Clicking SMPTE Test button on Maintenance Setup screen, the operator can perform SMPTE test
on the monitors.
Before using the equipment, verify that system operation is proper according to the check
items above. System that is not fully functional may result in delays in medical treatment.
CAUTION

Table of Contents

Other manuals for GE Brivo OEC 715

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Brivo OEC 715 and is the answer not in the manual?

GE Brivo OEC 715 Specifications

General IconGeneral
BrandGE
ModelBrivo OEC 715
CategoryMedical Equipment
LanguageEnglish

Related product manuals