Introduction and Safety
Page 1-14
Pulsed Low
Dose High
Level
Fluoro
8pps
kVp
77
77
76
68
69
70
110
110
mA
1.4
2
2.6
1.1
1.5
2.3
2
2
Dose Rate
(mGy/min)
2.53
3.61
4.56
0.73
1.05
1.68
7.77
4.86
Pulse
Normal
Fluoro
4pps
kVp
78
78
77
69
70
71
110
110
mA
2
2.8
3.9
1.5
2.2
3.3
4
4
Dose Rate
(mGy/min)
1.86
2.60
3.52
0.52
0.80
1.25
7.77
4.86
Pulsed
High Level
Fluoro
4pps
kVp
78
78
78
68
70
71
110
110
mA
3
4.2
6
2.3
3.3
4.5
6
6
Dose Rate
(mGy/min)
2.79
3.90
5.58
0.77
1.20
1.71
11.66
7.29
Pulsed Low
Dose 4pps
kVp
77
76
75
66
68
69
110
110
mA
0.9
1.2
1.7
0.6
0.9
1.4
2
2
Dose Rate
(mGy/min)
0.81
1.05
1.44
0.18
0.30
0.49
3.89
2.43
Pulsed Low
Dose High
Level
Fluoro
4pps
kVp
77
77
77
67
69
70
110
110
mA
1.3
1.8
2.6
1
1.5
2.1
2
2
Dose Rate
(mGy/min)
1.17
1.62
2.35
0.32
0.52
0.76
3.89
2.43
Pulse
Normal
Fluoro
2pps
kVp
78
78
77
68
70
71
110
110
mA
2
2.8
3.9
1.5
2.2
3.3
4
4
Dose Rate
(mGy/min)
0.93
1.30
1.76
0.25
0.40
0.63
3.89
2.43
Pulsed
High Level
Fluoro
2pps
kVp
78
78
78
68
70
71
110
110
mA
3
4.2
6
2.3
3.3
4.5
6
6
Dose Rate
(mGy/min)
1.39
1.95
2.79
0.38
0.60
0.85
5.83
3.64
Pulsed Low
kVp
77
76
75
66
68
69
110
110

Table of Contents