Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 1-23
It is applicable for Ed4.0 only
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of this
equipment, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Use Recommendations
This product complies with IEC 60601-1-2 EMC standard for medical devices and with radio frequency
emission requirements per CISPR11 Group1 Class A standard limits. The system is intended for use in
hospitals.
The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11
class B is normally required) this equipment might not offer adequate protection to
radio-frequency communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
Adhering to the distance separation recommendation table, between 150 kHz and 2.5 GHz, will reduce
disturbances recorded at the image level but may not eliminate all disturbances.
However, when installed and operated as specified herein, the system will maintain its essential
performance.
For example, a 1W mobile phone (800 MHz to 2.5 GHz carrier frequency) shall be put 2.3 meters apart from
the system (in order to avoid image interference risks).
The use of accessories, transducers, and cables other than those specified may result in degraded
ELECTROMAGNETIC COMPATIBILITY of the system.