Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 4-7
CAUTION
Only functions applicable to the mode selected are active. If any function is active, then
the corresponding indicator lights are illuminated. The normal auto fluoroscopy mode is
the default setting.
Image Orientation
Use these keys to rotate or reverse the image produced on the left WorkStation monitor. Image Rotation
and Image Reversal are useful for displaying the image in any orientation that you desire.
Image rotation and image reversal are not available to be used with swapped image and when image
annotation or image measurement has been applied.
Image Rotation
Press the left key to rotate the image counter-clockwise or press the right key to rotate the image
clockwise.
Image Reversal
Press the Image Reversal keys to change the orientation of the image displayed on the WorkStations left
monitor. Press the left key to reverse the image from left to right. Press the right key to invert the image
from top to bottom.