Usual Operations
Page 5-12
Pulsed Normal Fluoro
Auto/ Manual
kVp:40 to 110
mA: 0.1 to 4
Reduce X-Ray dose to
patient
Pulsed High Level
Fluoro
Auto/ Manual
kVp:40 to 110
mA: 0.2 to 12
High quality and low-dose
Imaging
Pulsed Low Dose
Auto/ Manual
kVp:40 to 110
mA: 0.1 to 2
+
Reduce X-Ray dose to
patient
Pulsed Low Dose
High Level
Auto/ Manual
kVp:40 to 110
mA: 0.1 to 6
+ +
High quality and low-dose
Imaging
Digital Spot
Auto
kVp:40 to 110
0.2 to 16 mA @ 200V,
220V, 230V, 240V;
0.2 to 10 mA @ 100V,
110V, 120V.
Short duration, high mA and
high quality single imaging
Low Dose Digital
Spot
Auto
kVp:40 to 110
0.1 to 8 mA @ 200V,
220V, 230V, 240V;
0.1 to 5 mA @ 100V,
110V, 120V.
+
Short duration, high mA and
high quality single imaging
with a lower dose
Film
Manual
kVp:40 to 110
20 mA @ 200V, 220V,
230V, 240V;
10 mA @ 100V, 110V,
120V.
Capture an X-Ray image on
film