Usual Operations
Page 5-20
Country/
Region
MODE
(Dose measured at 30 cm above the Image Intensifier)
Fluoro
kVp: 40 to 110
mA: 0.1 to 4
Pulsed Fluoro
kVp: 40 to 110
mA: 0.1 to 4
High Level
Fluoro
kVp: 40 to 110
mA: 0.2 to 12
Pulsed High
Level Fluoro
kVP:40 to 110
mA: 0.2 to 12
Digital Spot
kVp: 40 to 110
mA: 0.2 to 16
Film
kVp:40 to 110
mA: 20
mAs:1 to 80
/Macau
Hungary
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Iceland
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
India
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Indonesia
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Iraq
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Ireland
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Israel
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Italy
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Jamaica
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Japan
<50mGy/min
<50mGy/min
<50mGy/min
<50mGy/min
<2mGy
<20mGy
Jordan
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Kuwait
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Laos
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Latvia
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Lebanon
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy