Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 5-23
Country/
Region
MODE
(Dose measured at 30 cm above the Image Intensifier)
Fluoro
kVp: 40 to 110
mA: 0.1 to 4
Pulsed Fluoro
kVp: 40 to 110
mA: 0.1 to 4
High Level
Fluoro
kVp: 40 to 110
mA: 0.2 to 12
Pulsed High
Level Fluoro
kVP:40 to 110
mA: 0.2 to 12
Digital Spot
kVp: 40 to 110
mA: 0.2 to 16
Film
kVp:40 to 110
mA: 20
mAs:1 to 80
Taiwan
(Republic of
China)
<87 mGy/min
<87 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Thailand
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Trinidad &
Tobago
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Tunisia
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Turkey
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
United Arab
Emirates
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
USA
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
United
Kingdom
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Uruguay
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Venezuela
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Vietnam
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
The highest value for the Low Dose imaging doesn’t exceed 50% of the highest value for the Normal Fluoro.
The highest value for the Low Level Pulsed Fluoro doesn’t exceed 50% of the highest value for the Pulsed
Normal Fluoro.