Revision C 250cx Series Maternal/Fetal Monitor 5-9
2036947-001
Troubleshooting: Diagnostic Control Screen
Replacing the Stepper Motor
1. Slide new stepper motor into place.
2. Replace the two screws which attach the motor to the frame.
3. Re-connect the stepper motor harness cable to the Recorder Board at J4.
Printhead Adjustments
Refer to “Recorder Top View” on page 5-12 and “Recorder Right Side View”
on page 5-12 for identification of the adjustment points outlined in this section, and
to assist in clarifying adjustment procedures. These figures show top and right side
views of the 250cx Series Monitor strip chart recorder assembly removed from the
unit’s chassis.
Table 14 gives a description of each recorder adjustment. Adjustments C through F
can be performed without removing recorder from unit’s chassis; however, they
require removing the monitor’s top cover.
The printhead may require performing one or both of the following offset
adjustment procedures.
Vertical Offset Adjustment
To check vertical offset alignment, run at least two monitor self-tests by depressing
the front panel Test button. The first horizontal trace should be printed 0.490 ± 0.002
inch from the right hand paper guide. (Refer to figure “Vertical Offset
Adjustment Self-Test Trace” on page 5-13.) Perform the following steps if the
trace does not fall within the specified offset range.
1. Loosen the two set screws labeled A in Figure on page 5-12.
2. Adjust hex-head screw labeled C in Figure on page 5-12, until test trace falls
within specified tolerance.
3. Tighten the two set screws labeled A.
Horizontal Offset Adjustment
Adjust the horizontal offset if any of the following symptoms are detected. If
skewing occurs, check for other malfunctions. Noticeable skewing of a vertical line
printed on the strip chart paper is usually accompanied by one of the following
symptoms:
Paper consistently curls to one side.
Printing of unequal weight occurs along the vertical trace line, from one end to
the other.
Dots are missing along the vertical trace.
Printing is too light following printhead replacement.
If the Horizontal Offset Adjustment is required, perform the following steps:

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals