5-30 250cx Series Maternal/Fetal Monitor Revision C
2036947-001
Troubleshooting: FAQs
How do I change the default
tocodynamometer reference value?
1. Enter Install Options Screen 2. (Refer to “How do I access the service setup
/ install options screens?” on page 5-27.)
2. In Install Options Screen 2, rotate the Trim Knob to highlight the Default TOCO
Reference: setting on the eighth line from the top (excluding the Title).
3. The current setting is one of the following: 5, 10, 15, 20 or 25 mmHg.
4. Once the Default TOCO Reference: setting is highlighted, press the Trim Knob. The
current setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting. Select a value.
6. Once the desired value is set, press the Trim Knob again to save the value. The
current value setting stops blinking.
7. To exit the Service mode, follow the instructions at the top of Page 16.
How do I change the date? 1. Rotate the Trim Knob to highlight the Setup softkey at the bottom of the display,
below the menu bar.
2. Once the Setup softkey is highlighted, press the Trim Knob.
3. The display will change to the General Setup screen.
4. In the General Setup screen, rotate the Trim Knob to highlight one of the Date:
setting fields on the top right corner of the display. These fields are for day (DD),
month (MMM), and year (YYYY).
5. After the desired field is highlighted (selected), press the Trim Knob. The current
setting displays in blinking inverse video.
6. Now rotate the Trim Knob to change the current date parameter setting.
7. Once you set the desired value, press the Trim Knob again to save the value. The
current value setting stops blinking.
8. Repeat Step 4 through Step 7 for any other date parameters that need to be set.
9. To exit the Service mode, follow the instructions at the top of Page 16.
How do I change the time? 1. Rotate the Trim Knob to highlight the Setup softkey at the bottom of the display,
below the menu bar.
2. Once the Setup softkey is highlighted, press the Trim Knob.
3. The display changes to the General Setup screen.
4. In the General Setup screen, rotate the Trim Knob to highlight one of the Time:
setting fields on the top left corner of the display. These fields are for hours (HH),
minutes (MMM), and seconds (YYYY). Note that the seconds field cannot be
selected or set.
5. After the desired field is highlighted (selected), press the Trim Knob. The current
setting displays in blinking inverse video.
6. Now rotate the Trim Knob to change the current time parameter setting.
7. Once you set the desired value, press the Trim Knob again to save the value. The
current value setting stops blinking.
8. Repeat Step 4 through Step 7 for any other date parameters that need to be set.
9. To exit the Service mode, follow the instructions at the top of Page 16.
Question Answer

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD