Page 47 of 79 [EN] English – K0571 Issue 1
User Manual Druck DPI611
Note: If there is an electrical function already selected it
will automatically be disabled. The display will show a
screen message.
Figure 2-22 Switch Test
e. Example Procedure: Measure Voltage with Internal
Voltage Source
Figure 2-23 shows CH1 set-up to measure a DC voltage
(± 30 V) or DC mV (± 2000 mV) with Internal Voltage Source
(e.g. for use with Resistive bridge).
Note: Internal Voltage is provided by connecting to the
Vo terminals on the front of the DPI611 and enabling
Voltage (10V) or Millivolts (10V) as the electrical function.