DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
8 - 42 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-10 Freeze key assy
8-2-10-1 Tools
Common philips screwdrivers
8-2-10-2 Preparations
Shut down the system and switch off the main breaker.
8-2-10-3 Removal procedure
1.) Remove the Main Keyboard Assy. Refer to the 8-2-6 "Main Keyboard Assy" on page 8-36.
2.) Remove the 3 screws (1-3) and disconnect 1 cable. Refer to the figure below.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-10-4 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-79 Freeze Key in the main keyboard
Figure 8-80 Freeze Key connector
Table 8-11 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-10. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-7-5
Freeze key validation
connector
1
2
3