EasyManua.ls Logo

GE LOGIQ P6

GE LOGIQ P6
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
8 - 64 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-25-3 Removal procedure (cont’d)
8.) Unplug the J14 connector, and unscrew 2 screws. Pull out the APS II Assy carefully.
Note: AC power input connector should be passed through the square hole on shassis bottom.
Refer to the Figure 8-104.
9.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-25-4 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-104 Pull out the APS Assy
Table 8-26 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-25. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-9-5
APS II Assy PCB Function Validation
Procedure
Section 10-5-2
Functional Checks (See Also Chapter 4)
Section 10-5-5
Physical Inspection

Table of Contents

Other manuals for GE LOGIQ P6

Related product manuals