RSDetection
User’s Manual
Copyright © 2016 General Electric Company. All Rights Reserved S131-200-UM Rev A
Page 71
8.6.4 “A” Commands
Table 24 "A" Command Support
Sets the Dose Rate recording interval for the M command
AY4XXXXX<CR><LF>
Y – Unit address
XXXXX – Storage interval. Valid values are:
5, 10, 15, 30, 60, 120, 300, 600, 900
Restarts the Unit
AY6<CR><LF>
8.6.5 “P” Commands – Configuration Commands
Since all of the ‘P’ commands follow the same format, the format for each command will
not be described in the table. The command format is as follows.
#P Y PPP DDDD<CR>
Y – Unit address
PPP – Configuration parameter shown in Table 25.
DDDD – A variable length field used when setting the parameter value.
The database is used in the RSDetection system to store data, and the configuration
parameters described in the XML command section should be the preferred method of
configuration. These P commands are provided for legacy command support.
Table 25 "P" Command Support
Number of points in the queue. Always returns 20000
High alarm level. Returns 00000
Low alarm level. Returns 00000
Number of points in the queue. Always returns 20000
High alarm level. Returns 00000
Low alarm level. Returns 00000