restart the unit after this command is sent for the changes to
go into effect.
Sets or returns the RS=232 port baud rate. Valid values are:
300,600,1200,2400,9600,19200. User must restart the unit after
this command is sent for the changes to go into effect.
Parity for the RS-232 port. Valid values are: (E, O, N). User must
restart the unit after this command is sent for the changes to
go into effect.
Handshaking for the RS-232 port. Valid values are:
0 = No handshake, 1=RTS/CTS, > 1 delay in mSec. User must
restart the unit after this command is sent for the changes to
go into effect.
Sets or returns the RS=232 port baud rate. Valid values are:
300,600,1200,2400,9600,19200. User must restart the unit after
this command is sent for the changes to go into effect. Make
sure COM4 is available on the unit before sending this
command.
Parity for the RS-232 port. Valid values are: (E, O, N). User must
restart the unit after this command is sent for the changes to
go into effect. Make sure COM4 is available on the unit before
sending this command.
Handshaking for the RS-232 port. Valid values are:
0 = No handshake, 1=RTS/CTS, > 1 delay in mSec. User must
restart the unit after this command is sent for the changes to
go into effect. Make sure COM4 is available on the unit before
sending this command.
Set returns the serial port used by the display.
0 = No display, 1 = COM1, 2 – COM2, 4 = COM4.
External display units 0 = R/h, 1 = Sv/h.