2-2 Initial Setup
Entering the Program
Menu (cont.)
When entering information in an option window, press:
the [T] key to scroll down the list
the [S] key to scroll up the list
the [F2] key (Cancel) or the [ESC] key to close a window at any time
and return to Operate Mode without changing the data.
Note: If you enter an incorrect numeric value, press the [W] key to
erase the last digit entered.
Entering the Transducer
Parameters
To program the Transducer data, scroll to the Transducer option on
the Program Menu and press [ENTER] (see Figure 2-2 below).
Note: Refer to your Transducer Installation Guide for additional
information about transducers and configurations.
.
Figure 2-2: The Transducer Option Window
1. At the Type prompt, use the [W] and [X] keys to select the type of
transducer used in your installation (wetted or clamp-on). Press
[ENTER] to confirm the choice.
Note: Each choice made in the Transducer and Pipe menus will
determine the additional options that must be programmed.
Transducer
Pipe
Lining Path
Transducer/Pipe
OK
Cancel
Fluid
Type: Wetted
Clamp-on
Transducer
Special
Frequency
2.00
Tw
14
MHz
Wedge Ang
50
°
Wedge Tmp
25
°C
Wedge SS
1219.2
m/s
µs

Table of Contents