Initial Setup 2-11
August 2008
Entering Reynolds
Correction Data (cont.)
Entering a Single K/V Value
a. If you chose to enter a single value, press [ENTER] to open the
text box.
b. Use the numeric keys to enter the K/V value for your fluid (see
the Sound Speeds and Pipe Size Data booklet).
c. Press
[ENTER] to confirm your data.
Entering a Table of K/V Values
a. If you chose to enter a table of values, press [ENTER] to open the
data source box.
b. Press [F3] (OK) to confirm velocity as the source parameter.
c. At the Edit Table prompt, press [ENTER] to open the table of
data. Use the arrows keys to scroll to each table cell, and use
the numeric keys to enter the value for each cell.
d. When done, press [F3] (OK) to confirm the new table.
3. To return to Operate Mode, press [F3] (OK) to confirm your entries
or press [F2] (Cancel) to cancel your entries.
Entering a Calibration
Factor
Scroll to the “Calibration Factor” tab (see Figure 2-8 below), then
press [ENTER] and proceed as follows:
Note: The Calibration Factor is a multiplier used to match the
readings of the PT878 to another flow reference.
Figure 2-8: The Calibration Factor Window
Reynolds Correctio Inputs
Correction Factors
Calibration Factor
Calibration Factor
On
Off
Meter K-Factor
Data Source
K Factor
Single
Table
10
Edit Table
Cancel
OK