August 2008
2-10 Initial Setup
Entering Correction
Factors
To enter Reynolds Correction or Calibration Factor data, scroll to the
Correction Factors tab in the Program Menu and press [ENTER].
Entering Reynolds
Correction Data
Scroll to the “Reynolds Correction” tab (see Figure 2-7 below), then
press [ENTER] and proceed as follows:
Note: The Reynolds Correction factor is based on the kinematic
viscosity and the flow rate of the fluid. The default condition is
ON, and it should be left that way for all clamp-on transducer
applications and most wetted transducer applications.
Figure 2-7: The Reynolds Correction Window
1. To enable or disable Reynolds Correction, use the [
W] and [X]
keys to scroll to the appropriate radio button and press [ENTER].
Note: If you disable Reynolds Correction, you cannot enter any
other data in this window. Skip the remaining steps.
2. You must now choose to enter either a single K/V (kinematic
viscosity) value or a table of values. Use the [
W] and [X] keys to
scroll to the desired radio button and press [ENTER].
Proceed to the appropriate sub-section on the next page.
Reynolds Correctio
Inputs
Correction Factors
Cancel OK
Calibration Factor
Re Correction
On
Off
K/V (v)
m
3
/s
Data Source
K/V Factor
Single
Table
Edit Table