Measurement and Analysis
Vivid S5/Vivid S6 User Manual 285
R2424458-100 Rev. 2
measurement) is displayed and highlighted on the ECG.
A message appears.
2. To keep the current AVC setting, press
SET. To change the
AVC setting, use the trackball to display another frame and
press
SET.
If the AVC setting was changed, a Confirmation window is
displayed. Select one of the following options:
• Manual to accept the manual AVC setting.
• Event timing to discard the manual AVC setting (if for
example the AVC setting was not possible to assess
from the APLAX view). The AVC event timing
measurement will then be used.
Note: This choice is only visible if AVC event timing
measurement has been done.
• Auto to discard the manual AVC setting and use the
automatic AVC timing.
The Parametric systolic strain APLAX view is displayed
(Figure 7-17).
Figure 7-17: Parametric systolic strain APLAX view
Note: The image will not be saved unless Store is pressed.
Press Quad softkey to display a quad screen (Figure 7-18)
showing:
• 2D image with the ROI