Measurement and Analysis
Vivid S5/Vivid S6 User Manual 295
R2424458-100 Rev. 2
Editing the endocardial border tracing
1. If required, use the Left/Right Edge Shift softkey controls
to delineate separately the left or right portions of the
endocardial border visually as best as possible.
2. If required, move the trackball cursor over the border trace,
select an anchor point (red circle), press
SET and correct
the trace by dragging the anchor point to a new location
(Figure 7-27). The shape of the border trace will update
accordingly.
3. In case you wish to start over, press Redraw and
re-position the 3 points defining the basal and apex points.
Figure 7-26: Defining a ROI