Measurement and Analysis
298 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
The End Diastolic volume (EDV) and end Systolic Volume
(ESV) is calculated and shown above each frame.
The resulting EF calculation is displayed on top of the screen.
Tracking Validation
The tracking must be visually controlled and validated. If the
tracking results are visually correct press the red Approve
button. The button turns green and is labeled Approved. The
calculated values will be stored and later appear in the
worksheet.
If tracking needs correction there are several options:
If required, press Layout softkey to enlarge the ES and ED
frames and position them side-by-side (Figure 7-29).
You may manually select a different ED frame or ES frame
using the ED frame or ES frame softkeys.
You may edit any misaligned point on either ED or ES
frames. This is done by:
a. Placing the trackball cursor over the desired point.
b. Clicking over the red circle, it will turn to a square.
c. Pressing SET. The ESV, EDV and EF will be
recalculated accordingly.
d. Moving the square to re-align the border trace
(Figure 7-30) by:
Figure 7-30: Border Trace Edit
In case these results are incorrect you may go back to the
previous step by pressing the Recalc soft-button and
editing the endocardial border ("Tracing the endocardial