Measurement and Analysis
340 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Configuration of the Measurement result table
If AV CO calculation is not displayed in the Measurement result
table, follow the following procedure:
1. Press
CONFIG and select the category Meas/Text.
The Measurement menu sheet is displayed.
2. The AV CO calculation is based on Doppler AV Trace
measurement in the folder Aortic, check Doppler in the
Measurement menu sheet and select the folder Aortic.
A list of all available measurements and calculations for
the selected folder is displayed in the Measurement menu
sheet.
Note: Entries in green are calculated measurements.
3. In the Measurement menu sheet, double-click on the AV
Trace measurement.
A list of all available measurements and calculations for
the AV Trace measurement is displayed in the
Measurement menu sheet.
4. Check the box in front of AV CO.
The AV CO calculation will be displayed in the
Measurement result table.
Normal values
Normal values can be defined by the user for all parameters. A
Normal value can be either a range or a threshold. Normal
values entered are grouped by measurement category (e.g.
Cardiac, Pediatry etc.)
Normal values are displayed in the report.