Archiving
Vivid S5/Vivid S6 User Manual 447
R2424458-100 Rev. 2
Figure 9-32: Modifying/Creating the IP address
3. In the Properties window, enter:
• The DICOM server AE title. This entry is case sensitive
and must match exactly.
The DICOM server port
For some DICOM servers, the default Timeout setting
may be too low.
When configuring the DICOM storage device, the following
image settings should be entered in the Properties
window:
• Check DICOM SR if required (see below).
• Keep Reopen per image unchecked.
• Keep Allow raw data unchecked.
• Set Max Frame rate to 30.
• Keep Only Black and White unchecked.
• Set Compression to JPEG.
• Set Quality to 95.
• Check Allow multiframe.
DICOM SR
DICOM Structured Reporting (SR) is a standardized format for
medical results. Vivid S5/Vivid S6 and EchoPAC PC support
the specialized form for Adult Echo Ultrasound ("Supplement
72") for M&A results.
With the DICOM SR support, M&A for an exam can be sent at
the end of the exam or when exported from local archive. The
destination can be either a server on the network (Storage

Table of Contents