Archiving
448 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
SCP) or a removable media (DICOM Media) depending on the
DICOM dataflow selected.
"Supplement 72" does not support all M&A results from
Vivid S5/Vivid S6 and EchoPAC PC. "Supplement 72" limits the
information that is possible to send to the following:
• Publicly coded parameters, no pediatric or fetal cardiac or
unassigned measurement. Refer to the Vivid S5/Vivid S6
Reference manual for a complete list of supported
parameters.
• Basic modes: 2D, M-mode, Color Flow, PW and CW
Doppler.
• Publicly coded methods, not Modified Simpson or Bullet.
Refer to the Vivid S5/Vivid S6 Reference manual for a
complete list of supported methods.
• Basic derivations (Average, Last), no references between
the derived measurements and the ones they were made
from.
• Wall Motion Scoring: individual segment scores only
according to 16-segment model, no graded Hypokinesis
(only Hypokinesis is used).
Activating DICOM SR
DICOM SR must be activated for each DICOM device.
1. Press
CONFIG and log on as administrator.
2. Select the Connectivity category and Dataflow subgroup.
The Dataflow sheet is displayed.
3. Select the DICOM dataflow to configure in the Name
pull-down menu (see Figure 9-29).
4. Select a DICOM storage device in the Selected devices
pane and press Properties.
The Properties window for the selected DICOM stogare
device is displayed.