Presets and System setup
Vivid S5/Vivid S6 User Manual 619
R2424458-100 Rev. 2
WL-LA-DServ:
the local
database is not
searched, only
the DICOM
Modality
Worklist.
Worklist/LocalArchive-DI
COMServer/Int.HD
Modality Worklist local archive
DICOM server and local hard drive
Search in the DICOM Modality
Worklist, the patient found is copied
into local database. The patient
information and the examination
results are stored to the local the
database. Images are stored to a
DICOM Server and to an image
volume on the local hard drive.
Some of the measurements are stored
if DICOM SR is turned on (see
page 447)
Worklist/RemoteArchive-
DICOMServer/RemoteHD
Modality Worklist remote archive
DICOM server and remote hard drive
Search in the DICOM Modality
Worklist, the patient found is copied
into a remote database. The patient
information and examination results
are stored to a remote database.
Images are stored to a DICOM Server
and to an image network volume as
pure DICOM in both locations.
Some of the measurements are stored
if DICOM SR is turned on (see
page 447)
Worklist/Remote Archive
- Remote Storage
This dataflow is used in a network
environment that includes Vivid HL7
Gateway. The patient list in the
Search/Create Patient window is
coming from Vivid HL7 Gateway
through DICOM Modality Worklist. All
patient data and images are stored to
a server.
Dataflow Description