Safety
2–2 INS 61301607027 Rev A US
HOT SURFACES AND PIPES
WARNING
BURN HAZARD: Turn OFF and lock out/tag out the steam and hot water
supplies before servicing the sterilizer. Allow the steam to dissipate and let
the sterilizer cool down before touching any piping or surfaces.
There are hot surfaces in the service area. Avoid touching any piping
that could contain steam.
The sterilizer uses hot water and steam which have the potential of
causing burns or serious injuries. Wear personal protective equipment
suitable for hot water and steam.
LOCKOUT
Always keep the door to the service area locked.
No personnel should be in the service area while the sterilizer is
running.
Never bypass the door limit switches of the sterilizer.
When servicing the unit, the CONTROLS DISABLE/ENABLE switch
should be set to DISABLE. Remove the key and keep it with you.
OTHER CONSIDERATIONS
Leakage in the system must be repaired without delay.
The sterilizer must be kept clean to ensure optimum performance.
Do not hose down the outside of the sterilizer with water.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 600 Series and is the answer not in the manual?

Getinge 600 Series Specifications

General IconGeneral
BrandGetinge
Model600 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals