600 Series Steam Sterilizers
INS 61301607027 Rev A US 5–13
CONNECTING THE PLUMBING Make the following plumbing connections:
1. Connect the plumbing between the ¾ NPT pump discharge orifice and
the ¾ NPT inlet orifice at the sterilizer water supply strainer.
NOTE
The customer must provide a pressure gauge in the plumbing connection
between the pump discharge orifice and the sterilizer water strainer.
The customer must provide a shutoff valve in the water supply line to the
sterilizer.
2. Connect the plumbing between the 1¼ NPT pump suction orifice and
the water supply line.
NOTE
The pipe size must be 1½ in. OD minimum (the minimum flow rate must be
at least 5.1 gallons/minute).
FIGURE 5–11. BOOSTER PUMP PLUMBING CONNECTIONS
WS-0242
Suction Orice
(1¼ NPT)
Discharge Orice
(¾ NPT)
to Cold Water
Inlet on Sterilizer
(¾ NPT)
Pressure Gauge
and Shutoff Valve
(by customer)
CW
CW
Cold Water
See Roughing-In Drawing for
Utilities Requirements
Union
(recommended)