600 Series Steam Sterilizers
INS 61301607027 Rev A US 4–19
UTILITY REQUIREMENTS
The sterilizer requires steam, cold water and electrical supplies as specified
on the roughing-in drawing. Sterilizers with an integral boiler require a hot
water supply (instead of steam) and three-phase electrical supply as well.
FIGURE 4–19. UTILITY CONNECTIONS
Hot Water
Shutoff
Steam
Shutoff
Cold Water
Shutoff
Cold Water Supply
Steam Supply
(units without steam boiler)
Hot Water Supply
(units with steam boiler)
Steam Boiler
(option)
Booster Pump
(option)
115 or 230 Vac
Single-Phase
115 Vac
Integral
Transformer
(option)
UPS
(option)
UPS
(option)
Universal
Transformer
(option)
Mains
Disconnect
Circuit
Breaker
230 Vac
Voltages
> 380 Vac
Single-Phase
Power Source
Mains
Disconnect
Circuit
Breaker
Three-Phase
Power Source
Heater Voltage
A07027-BW
Power Box