Alar
m
Alarm text Description
ET06 A06 UNLOCK CS Limit switches on the soiled side do not register unlocked position within
6 sec of unlocking.
Possible action:
a. Check if the limit switch ”open door” is activated within 6
seconds after unlocking starts.
b. Check the function of the locking cylinders and limit switch.
ET07 A07 CLOSE CS Limit switches on the soiled side do not register a closed position within
20 seconds after closing. If the limit switch on the soiled side is not
deactivated within 8 seconds after opening. The closed position for the
limit switch is not registered while the waster-disinfector is running.
Possible action:
a. Check if the limit switch ”open door” is activated in 20 seconds
after the opening has started.
b. Check if the limit switch ”closed door” is deactivated in 8
seconds after the opening has started.
c. Check the function of the door motor and limit switch.
ET08 A08 OPEN CS Limit switches on the soiled side do not register the open position within
20 seconds after opening. If the limit switch on the soiled side is not
deactivated within 2 seconds after opening.
Possible action:
a. Check if the limit switch "opened door" is activated in 20
seconds after opening has started.
ET09 A09 LOCK CS Limit switches on the clean side do not register the locked position within
6 seconds of locking. The locking position for the limit switch is not
registered while the washer-disinfector is running.
Possible action:
a. Check if the limit switch ”locked door” is activated in 6 seconds
after the locking has started.
b. Check the function of the locking cylinder and limit switch.
ET10 A10 UNLOCK CS The limit switches on the clean side do not register the unlocked position
within 6 seconds after unlocking.
Possible action:
a. Check if the limit switch "unlocked door" is activated in 6
seconds after the unlocking has started.
b. Check the function of the locking cylinder and limit switch.
ET11 A11 CLOSE CS The limit switches on the clean side do not register the closed position
within 20 seconds after closing. If the limit switch on the clean side are
not deactivated within 8 seconds after opening.
The limit switch's closed position is not registered while the washer-
disinfector is running.
Possible action:
a. Check if the limit switch ”closed door” is activated in 20 seconds
after the unlocking has started.