OP-Panel selectable parameters
PRE RINSE STANDARD
VALUE (type)
PF00 P RINSE SELECTOR: value 0.1
Pre rinse selector (Main parameter for selecting or deselecting the whole
pre rinse.)
0 (P)
CT00 P RINSE FILL TYPE: value 1, 2, 3, 4, 5 (CW, HW, CW+HW, booster,
HCW)
Switch for selecting rinse water type (parameter for selecting type of water
to be used in the pre-rinse.
1=Cold water, 2=Hot water, 3=Mixed cold and hot water, 4=Filling from
booster tank, 5= High quality water.)
1 (P)
LI00 P RINSE TEMP SP-value 0-90°C
Set pre-rinse temperature.
0°C (P)
TI00 P RINSE TIME: value 0-60 min
Pre rinse time at selected pre rinse temperature.
1 min (P)
PF16 P RINSE GUARANTEE: value 0.1
Switch for guarantee for pre-rinse temperature. (0=pre-rinse time starts
directly whatever temperature achieved, 1=pre-rinse time starts when set
pre-rinse temperature reached)
1 (P)
Page 69 of 166