Overheating protection (only on electrically heated washer-
disinfectors)
The overheat protection sits inside the element. If the overheating protection is tripped, it must be
replaced with a new element.
This must only be done by authorized personnel.
The machine is connected to the electricity supply and some components are live.
1
001800
1. Heating element