Dose monitor calibration - desired volume (2.1.8.2.3.3)
DOSE MONITOR CALIBRATION
DOSE MONITOR TYPE UNIT FACTOR
DET PUMP 1 PC ml 0.130
DESIRED VOLUME: 200.0
MEASURED VOLUME: 180.0
49.1 C
CHAMBER TEMP
P1 HC SHORT 00:12:24
01 STANDBY 00:02:22
A34 I/O FAULT
1.202
DISCHARGE PRESS
128.1 C
DRYER TEMP
CANCEL CALC
001813
1
1. CALC
Use the CALC button to confirm the measured volume and to return to the menu to select the dose
monitor (2.1.6.2.3). Recalculate the factor.
Page 54 of 166