Getinge 9128 - Automatic Calibration Setting Hi

Getinge 9128
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Automatic calibration setting Hi (2.1.6.2.2.2)
49,1 C
WASH1 TEMP SP
P1 HC SHORT 00:12:24
01 STANDBY 00:02:22
A34 I/O FAULT
00:03:00
WASH1 TIME
128,1 C
DRYER TEMP
CANCEL
HIGH
SAVE
SCROLL
UP
SCROLL
DOWN
+/-
AUTOMATIC CALIBRATION
SENSOR GAIN OFFSET VALUE
00 CHAMBER TEMP 1,000 0,0 49,1
LOW REF 0.0 HIGH REF 0.0
000727
2
1
3
1. HIGH REF
2. HIGH SAVE
3. CANCEL
CANCEL opens the Select automatic sensor calibration (2.1.6.2.2).
Use the HIGH SAVE button to confirm the reference HIGH and open the screen where a calibration
protocol can be printed (2.1.6.2.2.3).
The values for increasing and compensation as recalculated.
Page 50 of 166