Getinge 9128 - Set the Cell Constant; Set the Damping

Getinge 9128
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. Press "E" on the transmitter.
2. Enter code 22 with the "+" and "-" buttons.
3. Press "E". "Setup 1" appears on the display.
4. Choose "Output" with the "+" and "-" buttons.
5. Press "E". "Sel. Type" appears on the bottom line.
6. Select "Lin" with the "+" and "-" buttons on the top row.
7. Press "E".
8. Select "4-20 mA" with the "+" and "-" buttons.
9. Press "E".
10. Choose a low value 0/4 mA = 0.000 uS/cm with the "+" and "-" buttons.
11. Press "E".
12. Select high value 20 mA = 500 uS/cm (5 mS/cm for transmitter in high range) with "+" and "-"
buttons.
13. Press "E".
14. Press "+" and "-" simultaneously to log out.
Set the cell constant
1. Press "E" on the transmitter.
2. Enter code 22 with the "+" and "-" buttons.
3. Press "E". "Setup 1" appears on the display.
4. Hold down "E" until "Cell constant" appears on the display.
5. Enter the cell constant with the "+" and "-" buttons for the relevant sensor. See the calibration
certificate that is attached to the machine.
6. Press "E".
7. Press "+" and "-" simultaneously to log out.
Set the damping
1. Press "E" on the transmitter.
2. Enter code 22 with the "+" and "-" buttons.
3. Press "E". "Setup 1" appears on the display.
4. Hold down "E" until "Damping" appears on the display.
5. Enter damping = 10 (25 for transmitter in high range) with the "+" and "-" buttons.
6. Press "E".
7. Press "+" and "-" simultaneously to log out.
Page 130 of 166