WASH2, NEUTRALIZING, CHEM DISINFEC. (Continue) STANDARD
VALUE (type)
LI33 FWASH DOSCOND SP: value 0-5000 µS/cm
Selected conductivity for detergent dosage of final wash. (It is very
important if this function is used that the conductivity sensor is properly
checked and calibrated, and recalibrated every second month, and that if
the detergent types are changed a new start up has to be performed.)
Affects the process if PF28 is 1 only.
1600 µS/cm (P)
PF26 FWASH SOAK: value 0.1
Final Wash soak selector. (0=no soak function, 1=soak function in 1-10
min after 1-10 min of washing). The soak function is used when you want
to pause the main pump somewhere during the ongoing wash, e.g. the
main pump, and everything else, stops after 1 min of washing and is kept
stopped for 1 min and the washing will go on for the rest of the selected
total washing time.
0 (P)
TI12 F WASH SOAK TIME: value 0-10 min
Final wash soak time set point. (The time everything is paused)
Affects the process if PF26 is 1 only.
1 min (P)
TI13 FWASH WAITSOAK: value 0-10 min
Final Wash wait for soak time set point. (The waiting time before everything
is paused)
Affects the process if PF26 is 1 only.
1 min (P)
PF18 FWASH GUARANTEE: value 0.1
Final Wash temperature guarantee selector. (0= final wash time starts
directly whatever temperature achieved, 1= final wash time starts at
selected final wash temperature set point achieved).
1 (P)
Page 73 of 166