Error LED (333 Pump)
The ERROR light may activate (steady red) for any of the following reasons:
A pump motor has ceased to function or has been asked to function in a way that is not
possible.
Communication has failed with a connected pump.
Pressure limits have been exceeded (high, File 21 and low, File 22). Any attempt to run a
file that does not exist will activate the error LED.
Restart after a power failure has attempted to start File 23. If the file does not exist, the
software will stop the pump and activate the ERROR LED.
Input #2 was closed, attempting to start File 23. If the file does not exist, the software will
stop the pump and activate the ERROR LED.
A maintenance counter limit has been reached. See Resetting Maintenance Logs on page 52.
Pump operation will cease when the maintenance counter limit is reached, even if a
method program is running.
The audit trail is saturated.
There is an invalid setting.
54
TROUBLESHOOTING | 333/334 HPLC PUMPS
ERROR LED (333 PUMP)