Error Messages (333 Pump)
Many types of messages are displayed on the screen; for instance, errors that occur during
the entry of a method program or configuration parameters. For this type of error, the
software displays messages that explain the required corrective action. However, there are
some critical error situations that can occur during or after a run.
Message Corrective Action
Pressure lower than ‘X’
Pressure higher than ‘Y’
Input 2 was activated
Check for leaks
Check configuration parameters
Check method parameters
Power restored after failure
The latest run could be invalid. Check the system
before a possible rerun of this file
Channel ‘Z’ failing
Failure may be caused by a blockage (tubing,
check valve, etc.)
Check the hydraulics, especially Channel ‘Z’
Check configuration parameters
Check method parameters
Communication fails with ‘N’
Check that pump ‘N’ is switched on
Check GSIOC ID and connections
File does not exist
Check that you are trying to run a valid file
Check that any linked files are valid
Maintenance limit is passed for ‘M’
Service item ‘M’ and reset the counter
See Resetting Maintenance Logs on page 52
Validation process has restored default
parameter(s)
Check all method program and configuration
parameters
RAM error contact your Gilson representative Contact your local Gilson representative
55
333/334 HPLC PUMPS | USER’S GUIDE
ERROR MESSAGES (333 PUMP)