75333/334 HPLC PUMPS | USERS GUIDE
FRONT PANEL CONTROL
The following options will be
available:
When power restored after
failure:
Start File 23
Continue from Same File
Stop
Input 2 close starts file [--]
Malfunction sounds beeper
alarm: YES / NO
Ensure that safety files started as the result of values entered here perform the desired
functions, such as stopping the pump. Files 21 through 24 are reserved for safety files. Do not
assign method programs to these file numbers. If you do not create the file, an error message
will display and the pump will stop.
File numbers 21 through 24 do not exist until you create the safety programs. Check that
these programs function correctly to ensure the desired level of protection.
4. Use the Up or Down arrows to navigate to each option. Press Change to cycle through
available selections for each option.
5. Press ENTER to confirm your selection.
When power restored after failure: [--]
Input 2 close starts le [--]
Malfunction sounds beeper alarm: YES / NO
Other Safety Functions Screen
Next
Figure 71
Other Safety Functions Screen

Table of Contents