83333/334 HPLC PUMPS | USER’S GUIDE
FRONT PANEL CONTROL
Looping Method Files
Method files can be looped to
repeat operations. From the Start
screen press Edit > File > “Select
File Number” > Edit to reach the
Edit Method screen.
Key in the value (1–999) for the
Number of loops line.
Press ENTER.
Linking Method Files
Method files can be linked to each
other to operate as a chain. From the
Start screen press Edit > File > “Select File Number” > Edit to reach the Edit Method screen.
1. Use the Down arrow key to navigate to the When Finished, Link to File field.
2. Press the Link To soft key.
3. Key in the file number.
4. Press ENTER.
Listing Events in Method Files
From the Start Screen press Edit > File > “Select File Number” > Edit to reach the Edit Method
screen.
1. Press the Events soft key.
2. Use the Up or Down arrows to navigate the list of events.
Parameters may be changed from the list of events. Use the Up or Down arrows to navigate
the list of events and use the Left and Right arrows to move the cursor from one parameter
field to another. After keying in each new value, press ENTER to confirm the value.
Navigation
● Press the Delete soft key to erase an event. Press Yes to confirm the deletion or No to return
to the previous screen.
● Press the Time/Flow soft key to toggle between the injection flow rate and duration view.
● Press the Change soft key to switch the state of an output.
● Press the Select soft key to edit an event.
Method file: [’X’]
Number of loops: [--]
When finished, link file to: [--]
Events Link To
Figure 79
Edit Method Screen