GDS-1000B Programming Manual
38
acquisition settings that precede the
raw data.
For example if the raw data for a point
is 001C (=28 decimal) then,
(28/25) x 0.5 = 0.56V
Example
:ACQuire1:MEMory?
Format,1.0B;Memory Length,10000;IntpDistance,0;
Trigger Address,4999;Trigger Level,1.160E+01;
Source,CH1;Vertical Units,V;Vertical Units
Div,0;Vertical Units Extend
Div,15;Label,ACK ;Probe Type,0;Probe
Ratio,1.000e+01;Vertical Scale,5.000e+00;Vertical
Position,-1.100e+01;Horizontal Units,S;Horizontal
Scale,5.000E-04;Horizontal Position,0.000E+00;
Horizontal Mode,Main;SincET Mode,Real
Time;Sampling Period,5.000e-07;Horizontal Old
Scale,5.000E-04;Horizontal Old Position,0.000E+00;
Firmware,V0.99b8;Waveform Data;
#520000………………..follows waveform block
data in hex…………………..
:ACQuire:FILTer:SOURce
Description
Returns the source of the filter.
Note
This command is only applicable when the Digital
Filter app is installed. See the Digital Filter app
user manual for details.
Syntax
:ACQuire:FILTer:SOURce {CH1|CH2|CH3|CH4|?}
Parameter/
Return parameter
CH1 ~ CH4
Source channel
Example
:ACQuire:FILTer:SOURce?
CH1
Sets the filter source to CH1.
Set
Query