COMMAND DETAILS
81
Example
:MEASure:SOURce1 CH1
:MEASure:FOVShoot?
1.27E+0
Selects Channel 1, and then measures the fall
overshoot.
:MEASure:FPReshoot
Description
Returns fall preshoot amplitude.
Syntax
:MEASure:FPReshoot{?}
Related
Commands
:MEASure:SOURce<X>
Returns
Returns the fall preshoot as <NR3>.
Return parameter
<NR3>
Returns the fall preshoot as a percentage.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:FPReshoot?
Selects Channel 1, and then measures the fall
preshoot.
:MEASure:FREQuency
Description
Returns the frequency value.
Syntax
:MEASure:FREQuency{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the frequency in Hz.
Chan Off
Indicates the source channel is not
activated.
Query