GDS-1000B Programming Manual
84
Return parameter
<NR3>
Returns the positive pulse width.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:PWIDth?
5.0E-6
Selects Channel 1, and then measures the positive
pulse width.
:MEASure:RISe
Description
Returns the first pulse rise time.
Syntax
:MEASure:RISe{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the rise time.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:RISe?
8.5E-6
Selects Channel 1, and then measures the rise time.
:MEASure:ROVShoot
Description
Returns the rising overshoot over the entire
waveform in percentage.
Syntax
:MEASure:ROVShoot{?}
Query
Query