COMMAND DETAILS
87
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:NPULSE?
4.000E+00
Selects Channel 1, and then measures the number
of negative pulses.
:MEASure:PEDGE
Description
Returns the number of positive edges.
Syntax
:MEASure:PEDGE{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the number of positive edges.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:PEDGE?
1.100E+01
Selects Channel 1, and then measures the number
of positive edges.
:MEASure:NEDGE
Description
Returns the number of negative edges.
Syntax
:MEASure:NEDGE{?}
Related
Commands
:MEASure:SOURce<X>
Query
Query