REMOTE CONTROL
145
Command
:LOAD[1|2]: RESistor?
Function
Queries the Load CR level.
Description
<NRf>
1-1000
Example
:LOAD2:RESistor?
Return the Load CR level for CH2.
Command
:LOAD:DISPlay {Low|Middle|High}
Function
In load mode, set the minimum voltage value that
can be displayed. The factory default is high. Below
this value, it will display --.----“.
Description
Low: 0.5VMiddle: 0.75VHigh:1V
Example
:LOAD:DISPlay Low
In load mode, the minimum voltage value that can
be displayed is 0.5V.
Command
:LOAD:DISPlay?
Function
Query the minimum voltage value that can be
displayed in load mode.
Example
:LOAD:DISPlay?
In load mode, query the minimum voltage value
that can be displayed. The return variables are low,
0.5V; middle, 0.75V; high, 1V.
Command
:DELAy[1|2]:CYCLEs {N|I}[,<value>]
Function
Sets the cycle number of Delay.
Description
<value> 1-99999
I represents an infinite loop. N represents a finite
loop. The cycle number is set by parameter
<value>.