REMOTE CONTROL
163
:SEQUence[1|2]:ENDState {OFF|LAST}
Sets terminate state of sequence.
{OFF|LAST} Output disabled (OFF): instrument
automatically disable output when output is
complete. The last group (LAST): instrument stays
at the output state of the last group when output is
complete.
:SEQUence2:ENDState LAST
Sets the terminate state of sequence as “the last
group”.
Queries the terminate state of the sequence. See
previous instruction.
:SEQUence2:ENDState?
Queries the CH2terminate state of the sequence.
Returns to OFF or LAST.
:SEQUence[1|2]:GROUPs <value>
Sets the output group number of sequence.
<value> the definition of output group number is
that the default voltage/current group number of
power output from each cycle ranges from 1 to
2048. All group number of sequence = output
group number*cycle number. Power will terminate
timer function after completing outputs of all
group numbers.
:SEQUence2:GROUPs 25
Sets the CH2 output group number of sequence as
25.