GPP-3060/6030 User Manual
164
Function
Queries the output group number of the sequence.
Example
:SEQUence2:GROUPs?
Queries the CH2 output group number of the
sequence. Returns to an integer between 1 to 2048.
Command
:SEQUence[1|2]:PARAmeter<No>,<volt>,<curr>,<
time>
Function
Sets the parameter of the specified group.
Description
< No > is the serial number of sequence parameter
of the specified group, integer, 0 to 2047.
<volt> is the voltage of sequence parameter of the
group. Unit is V, number, voltage range of the
active channel.
<curr> is the current of sequence parameter of the
group. Unit is A, number, current range of the
active channel.
<time> is the time of sequence parameter of the
group. Unit is s, number, 1s to 300s.
Example
:SEQUence2:PARAmeter 1,8,1,10
Sets the CH2 sequence parameter of the 1st group
as 8V,1A,10s.
Command
:SEQUence[1|2]:PARAmeter? <No>,< count>