FUNCTION MENU
121
4. Select Memory and select which memory
location to load for the selected step.
Settings loaded from the memory location will
be used for the selected step.
The same memory location can be used for
multiple steps.
Memory
M001 - M256
5. Set the Run setting for the step.
By default RUN is set to Skip.
The Auto setting will automatically start and go
onto the next step.
The Manual setting will wait for the user to
press Next[F2] before running the step.
Run
Skip, Auto, Manual
6. Choose the On-Time in seconds.
The on-time setting determines how long the
load is turned on for the selected step.
The on-time is defined as the total test time
minus the off-time.
On-Time
0.1 - 60 seconds
7. Choose the Off-Time in seconds.
The off-time setting determines how long the
load is turned off between the end of the
current step and the start of the next step.
The off-time is defined as the total test time
minus the on-time.
Off-Time
Off, 0.1 - 60 seconds
8. Choose the P/F-Time (pass/fail time) in
seconds.
The P/F-Time refers to the P/F delay time. This
delay time includes the 0.06 P/F start test time,
as shown in the timing diagram on page 119.