C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
• Install the analyzer indoors, in a non-hazardous environment.
• Install the analyzer in an environment that is protected from corrosive fluids.
• Install the analyzer in a clean, dry, well-ventilated and temperature-controlled location.
• Install the analyzer as near to the sampling point as possible.
• Do not install the analyzer in direct sunlight or near a heat source.
• Make sure that there is sufficient clearance to make plumbing and electrical connections.
• Make sure to leave sufficient space in front of the analyzer to open the analyzer door. Refer to
Analyzer dimensions.
• Make sure that the ambient conditions are within operating specifications. Refer to Specifications
on page 3.
Although the analyzer is not designed for use with flammable samples, some EZ-analyzers use
flammable reagents. If the analyzer uses flammable reagents, make sure to obey the safety
precautions that follow:
• Keep the analyzer away from heat, sparks and open flame.
• Do not eat, drink or smoke near the analyzer.
• Use a local exhaust ventilation system.
• Use spark and explosion-proof appliances and lighting system.
• Prevent electrostatic discharges. Refer to Electrostatic discharge (ESD) considerations
on page 14.
• Fully clean and dry the instrument before use.
• Wash hands before breaks and at the end of the working period.
• Remove contaminated clothing. Wash clothing before reuse.
• These fluids must be handled in accordance with local regulatory agency requirements on
permissible exposure limits.
5.2 Mechanical installation
5.2.1 Attach the instrument to a wall
W A R N I N G
Personal injury hazard. Make sure that the wall mounting is able to hold 4 times the weight of the
equipment.
W A R N I N G
Personal injury hazard. Instruments or components are heavy. Use assistance to install or move.
W A R N I N G
Personal injury hazard. The object is heavy. Make sure that the instrument is securely attached to a
wall, table or floor for a safe operation.
English 11