Error/Warning message Possible cause Solution
Analysis results are unstable Micropump is defective Make sure the reagents are dosed
correctly and there is no air in the
tubing.
Peristaltic pump is defective Make sure that the drain and
sample pump are operating
correctly.
Valve is defective Make sure that the valves (sample,
REF1, REF2, clean) are operating
correctly.
Stirrer is defective Examine if there is a magnetic
stirrer bar in the analysis vessel and
if the solution is stirred during
analysis.
Dispenser is defective Make sure that the dispenser piston
is filled with liquid and that there is
no air in the tubing.
The position of tubes in the analysis
vessel is not correct
Examine the position of the tubing in
the analysis vessel. Make sure that
the drain tubing is at the back side
of the analysis vessel and in the
ringlets. Other tubing should be
above the liquid level.
The reagents have expired. Prepare a new set of reagents when
the reagent bottles are empty.
Flush/Prime all tubing before a
measurement is started.
E-stop/ Reinitialize the dispenser! When Emergency stop is pushed,
the dispenser stops and must be
started again.
Examine the dispenser. Push F2 >
Dispenser to start the dispenser
again.
Sensor pH/mV error The pH or mV electrode are
defective or not connected.
Examine if the electrode is
connected correctly. Examine the
electrolyte level in the electrode,
refill if necessary.
Titration error The titration did not measure an EP
or the maximum amount of titration
solution was added without getting
an endpoint pH or mV.
Examine if sample solution goes
into the vessel.
Examine if the dispenser is filled.
Make sure that the sensing part
of the electrode is fully in the
sample.
Make sure that the electrode is
filled with electrolyte solution.
Examine the level of reagents
and titrant solution. Fill if
necessary.
Result alarm The measured result is too high or
lower than the set values in the
results (F5 > Software > Results >
Alarm).
Identify if the previous calibration
measured correctly (slope ok?).
Make sure that the sample
concentration is correct.
Identify if the vessel is clean.
Clean if necessary.
English 35

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach EZ Series and is the answer not in the manual?

Hach EZ Series Specifications

General IconGeneral
BrandHach
ModelEZ Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals