Table 20 Parameters and default settings (continued)
Parameter Default setting
Security settings
Enable security Disabled
Enable data logging Enabled
Password expires in # days 60
Remember last # passwords 3
Automatically log off after # minutes 30
Administrator ID Admin
Administrator password 123456
User's full name Blank
User ID Blank
User password Blank
Change password Enabled
Bottles change settings
TOC calibration:
Manually accept results Enabled
Return to online mode Disabled
Slope limit 15%
R2 limit 0.99
Conductivity calibration:
Manually accepts results Enabled
Return to online mode Disabled
Slope limit from previous 5%
Slope limit from factory calibration 10%
Use calibration resistor Enabled
TOC validation:
Return to online mode Disabled
System suitability:
Return to online mode Disabled
Bottle mode
TOC calibration setup Use default settings
TOC calibration custom settings All standards selected, 3 repetitions
Schedule bottle test Current time