3.2.4 Running Indicator
HY8200 series has operation indicator. When the electronic load is operating, ON key will
illuminated with red. Press ON key again to turn off the electronic load and the indicator will be
extinguished.
Note: Yellow wireframe represents the cursor position, which is the currently selected mode.
4.Measurement Settings
This chapter mainly covers the following contents:
Mode Setting and Measurement
Independent Short Current Test
Parameter Input and Operation Control
Alarm Prompts
Local/Remote
4.1 Mode Setting and Measurement
HY8200 series have 7 common test modes, which is CC, CV, CR, CP, dynamic, list and battery.
Press Mode key to enter mode interface, use direction key or rotaty knob to select the mode and
then press Enter key to enter the selected mode interface.
Mode Introduction
No matter how the input voltage changes, the electronic load always
consumes the constant current.
The electronic load maintains the input voltage as the set value by
changing the current consumption.
The electronic load is equals to a resistance, linear change the current
with the changing of input voltage to maintain a constant resistance